OUS will procure 2 and have the option to procure 1 (similar to operation 2) single door steam sterilisers (approx. 2 sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year. Sterilisation device/container dimensions: (L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm),(L60 mm x B 40 mm x H20 mm),(L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34,5 cm.
Frist
Fristen for mottak av tilbud var 2019-03-27.
Anskaffelsen ble publisert 2019-02-20.
Leverandører
Følgende leverandører er nevnt i tildelingsbeslutninger eller andre anskaffelsesdokumenter:
Kunngjøring av konkurranse (2019-02-20) Gjenstand Anskaffelsens omfang
Tittel: Autoclaves
Referansenummer: 75082-75167
Kort beskrivelse:
OUS will procure 2 and have the option to procure 1 (similar to operation 2) single door steam sterilisers (approx. 2 sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
Sterilisation device/container dimensions:
(L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm),(L60 mm x B 40 mm x H20 mm),(L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34,5 cm.
OUS will procure 2 and have the option to procure 1 (similar to operation 2) single door steam sterilisers (approx. 2 sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
Sterilisation device/container dimensions:
(L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm),(L60 mm x B 40 mm x H20 mm),(L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34,5 cm.
Metadata for kunngjøring
Originalspråk: engelsk 🗣️
Dokumenttype: Kunngjøring av konkurranse
Kontraktens art: Varer
Forskrift: Det europeiske økonomiske samarbeidsområdet (EØS), med deltakelse av GPA-land
Felles innkjøpsordliste (CPV)
Kode: Autoklaver📦 Utførelsessted
NUTS-region: Oslo
🏙️
Prosedyre
Prosedyretype: Åpen anbudskonkurranse
Tilbudstype: Innlevering for alle delkontrakter
Tildelingskriterier
Det mest økonomisk fordelaktige tilbudet
OUS will procure 2 and have the option to procure 1 (similar to operation 2) single door steam sterilisers (approx. 2 sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
OUS will procure 2 and have the option to procure 1 (similar to operation 2) single door steam sterilisers (approx. 2 sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
Sterilisation device/container dimensions:
(L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15…
… mm),(L60 mm x B 40 mm x H20 mm),(L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34,5 cm.
… mm), (L60 mm x B40 mm x H20 mm), (L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34,5 cm.
External measurement are preferably the same for both of the surgical departments but is a challenge due to the container size.
Operating theatre 1 D2.4052:
Maximum space is:
Depth: enter the outer measurement depth (must be able to take container L70 mm x B28 mm x H20 mm, L60 mm x B28 mm x H13 mm), (L54B x B26 mm x 15 mm)
Width: 900 mm — 1 000 mm
Height: can be a little higher than 1 970 mm
Operating theatre 2 D3.4023:
External measurements:
Depth: 1 020 mm (must be able to sterilise the sterilisation device (must be able to hold the sterilisation device: (L58 mm x B28 mm x H21 mm), (L62 mm x B42 mm x H20 mm)/(L60 mm x B20 mm x H20 mm)
Width: 900 mm — 950 mm
Height: can be a little higher than 1 970.
The steam steriliser delivered must be able to be transported into the named sterilisation room without building intervention (structural works). OUS will not perform any alteration of doors/transport routes in the corridors to get the sterilisers into the sterilisation room, due to patient activity.
The steam steriliser delivered must be able to be transported into the named sterilisation room without building intervention (structural works). OUS will not perform any alteration of doors/transport routes in the corridors to get the sterilisers into the sterilisation room, due to patient activity.
Transport route:
Door into the sterilisation room
Operating theatre 2: 87.5 cm x 203.5 cm
Operating theatre 1: 98 cm x 209 cm
One extra steam steriliser at Rikshospitalet [the National Hospital] needs to be replaced and we would like a price for 1 device as an option.
Varighet: 48 måneder
Beskrivelse av opsjoner:
1.1 Steam steriliser (single door) with integrated steam generator and RO system (for OUS) similar to operating theatre 2.
2.1 Service contracts:
Abidance of the option in Point 1 must be 4 years, with CPI changing after 2 years.
The acceptance period for the option in point 2 is until the end of the guarantee period + 3 years.
The Contracting Authority will have a unilateral right, but not a duty to take up the option.
Utførelsessted
Hovedsted eller utførelsessted: Oslo University Hospital.
Juridisk, økonomisk, finansiell og teknisk informasjon Vilkår for deltakelse
Egnethet til å utøve yrkesvirksomheten:
Tax and VAT certificate:
The certificates shall be issued on a certified form from the tax collector/tax office (www.skatteetaten.no). The certificates must be no more than six months from the deadline for receipt of tenders. Any arrears must be explained/justified.
Foreign tenderers must submit corresponding confirmation that government taxes and fees have been paid in the country where the tenderer has its registered place of business, in accordance with the respective country´s rules for such certificates.
Økonomisk og finansiell stilling:
Credit evaluation/rating, not older than 3 months, based on the last known accounting figures. The rating must be carried out by a credit rating company with a licence to conduct this service.
Creditworthiness with no requirement for collateral will be sufficient to meet the requirement.
If a tenderer has a justifiable cause to not present the documentation.
That the Contracting Authority has requested, he can document his economic and financial capacity by presenting other documents that the contracting authority deem to be appropriate. (PPR § 16-4 (2)).
Teknisk og faglig kapasitet:
Tenderers shall document a quality assurance system that documents the tenderer's organisation and ability for carrying out a good delivery. Enclose a short description (or example Table of Contents) of the tenderer's quality system/methods for quality assurance, and a copy of any certificates.
Tenderers shall document a quality assurance system that documents the tenderer's organisation and ability for carrying out a good delivery. Enclose a short description (or example Table of Contents) of the tenderer's quality system/methods for quality assurance, and a copy of any certificates.
Tenderers shall document the organisation's technical and professional qualifications. Enclose a detailed description of the tenderer´s organisation, both the sales and service organisation, including an organisation chart. State where the tenderer´s service organisation is located.
Tenderers shall document the organisation's technical and professional qualifications. Enclose a detailed description of the tenderer´s organisation, both the sales and service organisation, including an organisation chart. State where the tenderer´s service organisation is located.
If tenderers will use subcontractors: the tenderer must be able to send a declaration of commitment on request, before signing the contract, indicating that the tenderer will have the necessary resources available throughout the contract period. Describe in which areas it may be appropriate to use subcontractors. Manufacturers of equipment are not considered subcontractors.
If tenderers will use subcontractors: the tenderer must be able to send a declaration of commitment on request, before signing the contract, indicating that the tenderer will have the necessary resources available throughout the contract period. Describe in which areas it may be appropriate to use subcontractors. Manufacturers of equipment are not considered subcontractors.
Prosedyre
Rettslig grunnlag: 32014L0024
Tidspunkt for mottak av tilbud: 12:00
Språk som tilbud eller forespørsler om deltakelse kan sendes inn på: dansk 🗣️
norsk 🗣️ + 1 til
svensk 🗣️
Tilbudets gyldighetsperiode: 2019-10-28 📅
Dato for åpning av tilbud: 2019-03-27 📅
Tidspunkt for åpning av tilbud: 12:00
Sted: Oslo University Hospital.
Utfyllende informasjon Klageinstans
Navn: Oslo Universitetssykehus HF
Postadresse: Sognvannsveien 20
Poststed: Oslo
Postnummer: 0371
Land: Norge 🇳🇴
E-post: mtu-anbud@ous-hf.no📧
Kilde: OJS 2019/S 039-089127 (2019-02-20)
Kunngjøring om tildeling av kontrakt (2020-03-25) Gjenstand Anskaffelsens omfang
Kort beskrivelse:
Oslo University Hospital will procure two units and one option (as per operation 2) single door steam sterilisers (approx. two sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
Sterilisation device/container dimensions: (L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm), (L60 mm x B40 mm x H20 mm), (L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34.5 cm.
Oslo University Hospital will procure two units and one option (as per operation 2) single door steam sterilisers (approx. two sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
Sterilisation device/container dimensions: (L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm), (L60 mm x B40 mm x H20 mm), (L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34.5 cm.
Totalverdi for anskaffelsen: 2 556 163 NOK 💰
Metadata for kunngjøring
Dokumenttype: Kunngjøring om tildeling av kontrakt
Prosedyre
Tilbudstype: Ikke aktuelt
Referanse Datoer
Sendt dato: 2020-03-25 📅
Publiseringsdato: 2020-03-27 📅
Identifikatorer
Kunngjøringsnummer: 2020/S 062-149732
Refererer til kunngjøring: 2019/S 039-089127
OJ-S-utgave: 62
Gjenstand Anskaffelsens omfang
Kort beskrivelse:
Oslo University Hospital will procure two units and one option (as per operation 2) single door steam sterilisers (approx. two sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
Oslo University Hospital will procure two units and one option (as per operation 2) single door steam sterilisers (approx. two sterilisation devices) for sterilisation of invasive medical equipment, i.a. surgical instruments, thermostable endoscopic equipment, anaesthesia equipment and sterilisation of breast implants. Steam sterilisers must have steam generators and a water purification system for steam production. Factory testing documentation must be in place and be approved by OUS before delivery. The number of cycles is anticipated to be approx. 3-4 per day per autoclave/240 working days per year.
Sterilisation device/container dimensions: (L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm), (L60 mm x B40 mm x H20 mm), (L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34.5 cm.
Sterilisation device/container dimensions: (L58 x B28 x H21), (L60 x B28 x H13), (L54 mm x B26 mm x H15 mm), (L60 mm x B40 mm x H20 mm), (L70 mm x B28 mm x H20 mm) but the chamber must be dimension to and be able to run containing equipment with a width of 34.5 cm.
Depth: enter the outer measurement depth (must be able to take container L70 mm x B28 mm x H20 mm, L60 mm x B28 mm x H13 mm), (L54B x B26 mm x 15 mm).
Width: 900 mm-1 000 mm
Height: Can be a little higher than 1 970 mm
Depth: 1 020 mm (must be able to sterilise the sterilisation device (must be able to hold the sterilisation device: (L58 mm x B28 mm x H21 mm), (L62 mm x B42 mm x H20 mm)/L60 mm x B20 mm x H20 mm)
Width: 900 mm-950 mm
Height: can be a little higher than 1 970
Opr. 2 : 87.5 cm x 203.5 cm
Operating theatre 1 : 98 cm x 209 cm
One extra steam steriliser at Rikshospitalet [the National Hospital] needs to be replaced and we would like a price for one device as an option.
Beskrivelse av opsjoner:
1.1: Steam steriliser (single door) with integrated steam generator and RO system (for OUS) similar to operating theatre 2;
2.1: Service contracts
Abidance of the option in point 1 must be four years, with CPI changing after 2 years.
The contracting authority will have a unilateral right, but not a duty to take up the option.
Tildeling av kontrakt
Dato for kontraktinngåelse: 2019-06-18 📅
Navn: Thune Produkter AS
Postadresse: Postboks 4456
Poststed: Hamar
Postnummer: 2326
Land: Norge 🇳🇴
Telefon: +47 40003880📞
E-post: post@thuneprodukter.no📧
Land: Innlandet
🏙️
Totalverdi for anskaffelsen: 2 556 163 NOK 💰
Informasjon om tilbud
Antall mottatte tilbud: 2
Kilde: OJS 2020/S 062-149732 (2020-03-25)